The Waveguide Mach−Zender Interferometer as
Atrazine Sensor
Posted on 1998-02-14 - 00:00
Immunoanalytical techniques used in combination with
a highly sensitive Mach−Zender waveguide sensor give a
device that is capable of fast on-line monitoring of immunoreactions. The on-line monitoring is especially
attractive when the sensor can be used as an environmental
probe. It is demonstrated that low concentrations of
atrazine can be measured with a sensitivity around the
EC limit of 100 ng/L. These measurements are performed with an inhibition technique. In combination
with
“a dynamic slope method” one complete measurement
is done within 10 min. Sensor measurements show the
same sensitivity and errors as parallel ELISAs.