Point-of-Care
Diagnosis of Tuberculosis Using a Portable
Nucleic Acid Test with Distance-Based Readout
Posted on 2024-12-12 - 12:37
Rapid
and accurate diagnosis of tuberculosis (TB) infection
in
resource-limited settings is critically needed to stop the spread
of the disease but remains difficult to achieve. Herein, we report
a fast, inexpensive nucleic acid test with distance-based readout
(FINDR) for TB. Based on the unique chromatographic behavior of DNA
intercalating dye on unmodified cellulose paper, our FINDR platform
converted the amplicons of loop-mediated isothermal amplification
(LAMP) into the migration distance on a paper-based analytical device.
A suite of innovations were further introduced to enhance the accessibility
of FINDR, including (1) screening optimal LAMP primers for distance-based
readout; (2) developing a chip-on-cover design capable of streamlining
LAMP amplification and distance-based detection in a fully sealed
FINDR chip with a simple hand-based swirling for liquid transfer;
and (3) integrating FINDR with an upstream portable pipetting-free
DNA extractor and a downstream smartphone app to simplify sample processing
and data interpretation. With these innovations, FINDR demonstrated
a high clinical sensitivity (96.6%) and specificity (100%) upon validation
against clinical sputum samples. Successful clinical validation was
also achieved when FINDR was deployed as a point-of-care test for
detecting TB under resource-limited conditions.
CITE THIS COLLECTION
DataCiteDataCite
No result found
Dong, Tianyu; Qin, Liwen; Wang, Zhiyin; Fan, Chen; Shen, Chenlan; Feng, Pin; et al. (2024). Point-of-Care
Diagnosis of Tuberculosis Using a Portable