Photonic Lab-on-a-Chip for Rapid Cytokine Detection
Version 2 2016-08-22, 13:32Version 2 2016-08-22, 13:32
Version 1 2016-07-07, 17:19Version 1 2016-07-07, 17:19
Posted on 2016-06-21 - 00:00
A photonic
lab-on-a-chip (PhLoC) was developed for rapid determination
of interleukin-2 (IL-2) levels secreted by lymphocytes using measurement
of optical absorbance. Optical and microfluidic components, including
lenses, mirrors, and flow channels, were monolithically created using
poly(dimethylsiloxane) (PDMS). Sample and reagent solutions were processed
in the form of liquid columns separated by air. A flow channel structure
with an air bypass was used to release air between solution columns
and to exchange solutions in the measuring chamber, avoiding the intrusion
of air into the measuring chamber and the denaturation of the capture
antibodies. The concentration of IL-2 was measured by enzyme-linked
immunosorbent assay (ELISA) in the PhLoC. The total time required
for the ELISA was less than 30 min. Clear concentration dependence
was confirmed for IL-2 concentrations ranging from 50 to 1000 pg/mL.
Thus, the concentration of IL-2 secreted by lymphocytes following
activation by concanavalin A (Con A) could be measured using the PhLoC.