Increasing the Multiplexing
Capacity of TMTs Using
Reporter Ion Isotopologues with Isobaric Masses
Posted on 2012-09-04 - 00:00
Quantitative mass spectrometry methods offer near-comprehensive
proteome coverage; however, these methods still suffer with regards
to sample throughput. Multiplex quantitation via isobaric chemical
tags (e.g., TMT and iTRAQ) provides an avenue for mass spectrometry-based
proteome quantitation experiments to move away from simple binary
comparisons and toward greater parallelization. Herein, we demonstrate
a straightforward method for immediately expanding the throughput
of the TMT isobaric reagents from 6-plex to 8-plex. This method is
based upon our ability to resolve the isotopic shift that results
from substituting a 15N for a 13C. In an accommodation
to the preferred fragmentation pathways of ETD, the TMT-127 and -129
reagents were recently modified such that a 13C was exchanged
for a 15N. As a result of this substitution, the new TMT
reporter ions are 6.32 mDa lighter. Even though the mass difference
between these reporter ion isotopologues is incredibly small, modern
high-resolution and mass accuracy analyzers can resolve these ions.
On the basis of our ability to resolve and accurately measure the
relative intensity of these isobaric reporter ions, we demonstrate
that we are able to quantify across eight samples simultaneously by
combining the 13C- and 15N-containing reporter
ions. Considering the structure of the TMT reporter ion, we believe
this work serves as a blueprint for expanding the multiplexing capacity
of the TMT reagents to at least 10-plex and possibly up to 18-plex.