ac1c01469_si_004.mp4 (2.25 MB)
Download fileAutonomous Internal Reflux of Magnetic Nanoparticle Chains in a Flow Channel for Efficient Detection of Waterborne Bacteria
media
posted on 2021-09-03, 02:43 authored by Hyeonjeong Lee, Hyunsoo Han, Sangmin JeonHerein, we developed a novel method
for the efficient capture of
waterborne bacteria by creating an autonomous internal reflux of the
magnetic nanoparticle chains (MNCs) inside a flow channel. A glass
tube containing positively charged polyethyleneimine-coated MNCs (PEI-MNCs)
was placed at the center of a Halbach ring, generating a strong and
uniform magnetic field inside the ring. When a bacteria-spiked solution
was injected into the tube, the target bacteria bound to the PEI-MNCs
via an electrostatic interaction remained in the tube, whereas the
unbound bacteria left the tube. Some PEI-MNC–bacteria complexes
left the glass tube at high flow rates because of the drag force,
which reduced the capture efficiency of the device. The loss of the
PEI-MNC–bacteria complexes at high flow rates was suppressed
by placing a k0 ring behind the Halbach
ring. The k0 ring was used to apply a
magnetic force in the opposite direction of the solution flow and
create an autonomous reflux of the PEI-MNCs inside the glass tube,
reducing their loss and increasing their capture efficiency. The capture
efficiency of Escherichia coli O157
was determined based on the cell count to be greater than 90% at a
flow rate of 15 mL/min. E. coli O157
was detected using quantitative polymerase chain reaction, and the
limits of detection were 2 and 3 cfu/mL in deionized water and river
water, respectively.
History
Usage metrics
Categories
Keywords
electrostatic interaction remainedmagnetic nanoparticle chainsunbound bacteria lefttarget bacteria boundhigh flow ratesescherichia coli o157autonomous internal reflux0 subwaterborne bacteria hereincoli o157autonomous refluxwaterborne bacteriamagnetic forcek >< subsolution flowflow rateflow channelspiked solutionriver wateropposite directionnovel methodefficient capturedrag forcedetermined baseddeionized watercell countcapture efficiency3 cfu