posted on 01.07.2007, 00:00 authored by Reto Thürer, Tamás Vigassy, Martina Hirayama, Joseph Wang, Eric Bakker, Ernö PretschPotentiometric sensors based on polymer membrane
electrodes, if properly optimized, are useful for measurements at trace levels. The expected independence of the
electrochemical signal of the sample size makes them
extremely attractive for measurements in small volumes.
Here, we report on electrodes for the potentiometric
detection of cadmium ions that reach a detection limit of
6 nM and utilize a Na+-selective electrode as pseudoreference in order to facilitate measurements in 150-μL
samples. A potentiometric immunoassay of mouse IgG is
performed via CdSe quantum dot labels on a secondary
antibody according to a sandwich immunoassay protocol
in a microtiter plate format. The CdSe quantum dots are
found to be easily dissolved/oxidized in a matter of
minutes with hydrogen peroxide, allowing us to maintain
the pH at a near-neutral value. The potentiometric protein
immunoassay exhibits a log−linear response ranging from
0.15 to 4.0 pmol of IgG, with a detection limit of <10
fmol in 150-μL sample wells.