posted on 2025-03-27, 15:19authored bySultan Mahmud, Vikum K. Dewasurendra, Chandrima Banerjee, Pedram Tavadze, Mst Nigar Sultana, Mohammad A. Rahman, Sohag Ahmed, Peng Li, Matthew B. Johnson, Stephen J. Valentine
Capillary vibrating sharp-edge spray ionization (cVSSI)
has been
used to study the effects of applied voltage and mass spectrometer
heated inlet transfer tube temperature on DNA triplex ion production
for native mass spectrometry (MS) samples. Overall, medium applied
voltage (−900 to −1000 V) results in better ion production
of the desired triplex ions (Tri) (i.e., those without cation adducts
such as NH4+, Na+, and K+); mass spectral peak intensities for the [Tri]8–, [Tri]9–, and [Tri]10– ions
increase by ∼70, ∼260, and ∼125 fold, respectively,
compared to higher voltages (−1100 to −1500 V). The
latter voltages result in increased triplex adduct ion (Tri + ad)
formation; for the 8–, 9–, and 10– charge states;
the ratios of Tri to Tri+ad ion abundances increase by ∼6 fold
for the lower voltage. By capillary inlet temperatures of 300 to 400
°C, Tri ion abundances reach maximum values of 6.1 × 105 ([Tri]8–), 2.9 × 106 ([Tri]9–), and 6.4 × 105 ([Tri]10–). Ion abundances for the respective species decrease by ∼4,
∼14, and ∼190 fold at a heated inlet transfer tube temperature
of 450 °C. The abundances for Tri+ad ions species generally follow
a similar trend as a function of heated inlet transfer tube temperature
with the exception that maximum values are obtained at 250 °C.
The abundances for DNA triplex fragment ions (Tri-fr) reach maximum
values at 400 °C resulting from excessive, in-source ion activation.
From these studies, the optimal capillary MS inlet temperature for
production of large oligonucleotides by cVSSI is 300 to 350 °C
and the applied voltage should be maintained at ∼ −900
V. These studies lay the foundation for native MS of large oligonucleotide
species in negative-ion mode exploiting the sensitivity enhancements
of cVSSI.