Inlet Ionization: A New Highly Sensitive Approach for Liquid Chromatography/Mass Spectrometry of Small and Large Molecules

Download (124.32 kB)
journal contribution
posted on 2011-10-15, 00:00 authored by Vincent S. Pagnotti, Ellen D. Inutan, Darrell D. Marshall, Charles N. McEwen, Sarah Trimpin
Inlet ionization is a new approach for ionizing both small and large molecules in solids or liquid solvents with high sensitivity. The utility of solvent based inlet ionization mass spectrometry (MS) as a method for analysis of volatile and nonvolatile compounds eluting from a liquid chromatography (LC) column is demonstrated. This new LC/MS approach uses reverse phase solvent systems common to electrospray ionization MS. The first LC/MS analyses using this novel approach produced sharp chromatographic peaks and good quality full mass range mass spectra for over 25 peptides from injection of only 1 pmol of a tryptic digest of bovine serum albumin using an eluent flow rate of 55 μL min–1. Similarly, full acquisition LC/MS/MS of the MH+ ion of the drug clozapine, using the same solvent flow rate, produced a signal-to-noise ratio of 54 for the major fragment ion with injection of only 1 μL of a 2 ppb solution. LC/MS results were acquired on two different manufacturer’s mass spectrometers using a Waters Corporation NanoAcquity liquid chromatograph.