High-Throughput Method
Development for Sensitive,
Accurate, and Reproducible Quantification of Therapeutic Monoclonal
Antibodies in Tissues Using Orthogonal Array Optimization and Nano
Liquid Chromatography/Selected Reaction Monitoring Mass Spectrometry
posted on 2016-02-20, 21:45authored byXiaotao Duan, Lubna Abuqayyas, Lipeng Dai, Joseph P. Balthasar, Jun Qu
Although liquid chromatography/mass spectrometry using
selected
reaction monitoring (LC/SRM-MS) holds great promise for targeted protein
analysis, quantification of therapeutic monoclonal antibody (mAb)
in tissues represents a daunting challenge due to the extremely low
tissue levels, complexity of tissue matrixes, and the absence of an
efficient strategy to develop an optimal LC/SRM-MS method. Here we
describe a high-throughput, streamlined strategy for the development
of sensitive, selective, and reliable quantitative methods of mAb
in tissue matrixes. A sensitive nano-LC/nanospray-MS method was employed
to achieve a low lower limit of quantification (LOQ). For selection
of signature peptides (SP), the SP candidates were identified by a
high-resolution Orbitrap and then optimal SRM conditions for each
candidate were obtained using a high-throughput, on-the-fly orthogonal
array optimization (OAO) strategy, which is capable of optimizing
a large set of SP candidates within a single nano-LC/SRM-MS run. Using
the optimized conditions, the candidates were experimentally evaluated
for both sensitivity and stability in the target matrixes, and SP
selection was based on the results of the evaluation. Two unique SP,
respectively from the light and heavy chain, were chosen for quantification
of each mAb. The use of two SP improves the quantitative reliability
by gauging possible degradation/modification of the mAb. Standard
mAb proteins with verified purities were utilized for calibration
curves, to prevent the quantitative biases that may otherwise occur
when synthesized peptides were used as calibrators. We showed a proof
of concept by rapidly developing sensitive nano-LC/SRM-MS methods
for quantifying two mAb (8c2 and cT84.66) in multiple preclinical
tissues. High sensitivity was achieved for both mAb with LOQ ranged
from 0.156 to 0.312 μg/g across different tissues, and the overall
procedure showed a wide dynamic range (≥500-fold) and good
accuracy [relative error (RE) < 18.8%] and precision [interbatch
relative standard deviation (RSD) < 18.1%, intrabatch RSD <
17.2%]. The quantitative method was applied to a comprehensive investigation
of the steady-state tissue distribution of 8c2 in wild-type mice versus
those deficient in FcRn α-chain, FcγIIb, and FcγRI/FcγRIII,
following a chronic dosing regimen. This work represents the first
extensive quantification of mAb in tissues by an LC/MS-based method.