se8b01277_si_002.zip (810.46 kB)

Extraction, Enrichment, and in situ Electrochemical Detection on Lab-on-a-Disc: Monitoring the Production of a Bacterial Secondary Metabolite

Download (810.46 kB)
dataset
posted on 11.12.2018 by Sune Zoëga Andreasen, Kuldeep Sanger, Christian Bille Jendresen, Alex Toftgaard Nielsen, Jenny Emnéus, Anja Boisen, Kinga Zór
Development of microsystems, which enable “sample-to-answer” detection from real samples, is often challenging. We present the first integration of supported liquid membrane extraction combined with electrochemical detection on a centrifugal fluidic platform. The developed lab-on-a-disc (LoD) system enabled the separation, enrichment, and subsequent electrochemical detection of the target analyte from a complex sample mixture. As a case study, we quantified the amount of a dietary supplement and pharmaceutical precursor, p-coumaric acid, from bacterial growth media at different time points during production. The assay, extraction, and detection, performed on the LoD device, proved to be a low cost and environmentally friendly approach, requiring only a few tens of microliters of organic solvent and enabled detection in a 3 μL volume. In addition, the data obtained from the centrifugal platform showed a good correlation with data obtained from the high performance liquid chromatography analysis.

History

Exports